CVE-2014-5406

Published Jul 6, 2015

Last updated 9 years ago

Overview

Description
The Hospira LifeCare PCA Infusion System before 7.0 does not validate network traffic associated with sending a (1) drug library, (2) software update, or (3) configuration change, which allows remote attackers to modify settings or medication data via packets on the (a) TELNET, (b) HTTP, (c) HTTPS, or (d) UPNP port. NOTE: this issue might overlap CVE-2015-3459.
Source
ics-cert@hq.dhs.gov
NVD status
Analyzed

Risk scores

CVSS 2.0

Type
Primary
Base score
9.3
Impact score
10
Exploitability score
8.6
Vector string
AV:N/AC:M/Au:N/C:C/I:C/A:C

Weaknesses

nvd@nist.gov
CWE-345

Social media

Hype score
Not currently trending

Configurations