CVE-2014-5434

Published Mar 26, 2019

Last updated 5 years ago

Overview

Description
Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 has a default account with hard-coded credentials used with the FTP protocol. Baxter asserts no files can be transferred to or from the WBM using this account. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.
Source
ics-cert@hq.dhs.gov
NVD status
Modified

Risk scores

CVSS 3.0

Type
Primary
Base score
9.8
Impact score
5.9
Exploitability score
3.9
Vector string
CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Severity
CRITICAL

CVSS 2.0

Type
Primary
Base score
5
Impact score
2.9
Exploitability score
10
Vector string
AV:N/AC:L/Au:N/C:P/I:N/A:N

Weaknesses

nvd@nist.gov
CWE-798
ics-cert@hq.dhs.gov
CWE-259

Social media

Hype score
Not currently trending

Configurations