CVE-2015-3952

Published Mar 25, 2019

Last updated 5 years ago

Overview

Description
Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
Source
ics-cert@hq.dhs.gov
NVD status
Modified

Risk scores

CVSS 3.0

Type
Primary
Base score
7.5
Impact score
3.6
Exploitability score
3.9
Vector string
CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
Severity
HIGH

CVSS 2.0

Type
Primary
Base score
5
Impact score
2.9
Exploitability score
10
Vector string
AV:N/AC:L/Au:N/C:P/I:N/A:N

Weaknesses

nvd@nist.gov
CWE-200
ics-cert@hq.dhs.gov
CWE-312

Social media

Hype score
Not currently trending

Configurations