CVE-2017-12712

Published Apr 25, 2018

Last updated 5 years ago

Overview

Description
The authentication algorithm in Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017, which involves an authentication key and time stamp, can be compromised or bypassed, which may allow a nearby attacker to issue unauthorized commands to the pacemaker via RF communications. CVSS v3 base score: 7.5, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.
Source
ics-cert@hq.dhs.gov
NVD status
Modified

Risk scores

CVSS 3.0

Type
Primary
Base score
8.8
Impact score
5.9
Exploitability score
2.8
Vector string
CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Severity
HIGH

CVSS 2.0

Type
Primary
Base score
8.3
Impact score
10
Exploitability score
6.5
Vector string
AV:A/AC:L/Au:N/C:C/I:C/A:C