CVE-2017-12718

Published Feb 15, 2018

Last updated 5 years ago

Overview

Description
A Classic Buffer Overflow issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. A third-party component used in the pump does not verify input buffer size prior to copying, leading to a buffer overflow, allowing remote code execution on the target device. The pump receives the potentially malicious input infrequently and under certain conditions, increasing the difficulty of exploitation.
Source
ics-cert@hq.dhs.gov
NVD status
Modified

Risk scores

CVSS 3.0

Type
Primary
Base score
8.1
Impact score
5.9
Exploitability score
2.2
Vector string
CVSS:3.0/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H
Severity
HIGH

CVSS 2.0

Type
Primary
Base score
6.8
Impact score
6.4
Exploitability score
8.6
Vector string
AV:N/AC:M/Au:N/C:P/I:P/A:P

Weaknesses

nvd@nist.gov
CWE-119
ics-cert@hq.dhs.gov
CWE-120

Social media

Hype score
Not currently trending

Configurations