CVE-2017-14018

Published Dec 5, 2017

Last updated 5 years ago

Overview

Description
An improper authentication issue was discovered in Johnson & Johnson Ethicon Endo-Surgery Generator Gen11, all versions released before November 29, 2017. The security authentication mechanism used between the Ethicon Endo-Surgery Generator Gen11 and single-patient use products can be bypassed, allowing for unauthorized devices to be connected to the generator, which could result in a loss of integrity or availability.
Source
ics-cert@hq.dhs.gov
NVD status
Modified

Risk scores

CVSS 3.0

Type
Primary
Base score
4.8
Impact score
4.2
Exploitability score
0.5
Vector string
CVSS:3.0/AV:P/AC:H/PR:N/UI:N/S:U/C:N/I:H/A:L
Severity
MEDIUM

CVSS 2.0

Type
Primary
Base score
3.3
Impact score
4.9
Exploitability score
3.4
Vector string
AV:L/AC:M/Au:N/C:N/I:P/A:P

Weaknesses

nvd@nist.gov
CWE-287
ics-cert@hq.dhs.gov
CWE-287

Social media

Hype score
Not currently trending

Configurations