CVE-2017-9658

Published Apr 30, 2018

Last updated 5 years ago

Overview

Description
Certain 802.11 network management messages have been determined to invoke wireless access point blacklisting security defenses when not required, which can necessitate intervention by hospital staff to reset the device and reestablish a network connection to the Wi-Fi access point. During this state, the Philips IntelliVue MX40 Version B.06.18 can either connect to an alternative access point within signal range for association to a central monitoring station, or it can remain in local monitoring mode until the device is reset by hospital staff. CVSS v3 base score: 6.5, CVSS vector string: AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H. Philips has released software update, Version B.06.18, to fix the improper cleanup on thrown exception vulnerability, and implement mitigations to reduce the risk associated with the improper handling of exceptional conditions vulnerability. The software update implements messaging and alarming on the MX40 and at the central monitoring station, when the MX40 disconnects from the access point.
Source
ics-cert@hq.dhs.gov
NVD status
Modified

Risk scores

CVSS 3.0

Type
Primary
Base score
6.5
Impact score
3.6
Exploitability score
2.8
Vector string
CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
Severity
MEDIUM

CVSS 2.0

Type
Primary
Base score
6.1
Impact score
6.9
Exploitability score
6.5
Vector string
AV:A/AC:L/Au:N/C:N/I:N/A:C

Weaknesses

nvd@nist.gov
CWE-755
ics-cert@hq.dhs.gov
CWE-755