CVE-2018-10601

Published Jun 5, 2018

Last updated 3 days ago

Overview

Description
IntelliVue Patient Monitors MP Series (including MP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M, IntelliVue Patient Monitors MX (MX400-550) Rev J-M and (X3/MX100 for Rev M only), and Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50 with software Revisions F.0, G.0 and J.3 have a vulnerability that exposes an "echo" service, in which an attacker-sent buffer to an attacker-chosen device address within the same subnet is copied to the stack with no boundary checks, hence resulting in stack overflow.
Source
ics-cert@hq.dhs.gov
NVD status
Modified

Risk scores

CVSS 3.1

Type
Primary
Base score
8.2
Impact score
6
Exploitability score
1.6
Vector string
CVSS:3.1/AV:A/AC:H/PR:N/UI:N/S:C/C:H/I:L/A:H
Severity
HIGH

CVSS 2.0

Type
Primary
Base score
5.4
Impact score
6.4
Exploitability score
5.5
Vector string
AV:A/AC:M/Au:N/C:P/I:P/A:P

Weaknesses

ics-cert@hq.dhs.gov
CWE-121
nvd@nist.gov
CWE-787

Social media

Hype score
Not currently trending

Configurations