CVE-2018-14781

Published Aug 13, 2018

Last updated 5 years ago

Overview

Description
Medtronic MMT 508 MiniMed insulin pump, 522 / MMT - 722 Paradigm REAL-TIME, 523 / MMT - 723 Paradigm Revel, 523K / MMT - 723K Paradigm Revel, and 551 / MMT - 751 MiniMed 530G The models identified above, when paired with a remote controller and having the "easy bolus" and "remote bolus" options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery.
Source
ics-cert@hq.dhs.gov
NVD status
Modified

Risk scores

CVSS 3.0

Type
Primary
Base score
5.3
Impact score
3.6
Exploitability score
1.6
Vector string
CVSS:3.0/AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:H/A:N
Severity
MEDIUM

CVSS 2.0

Type
Primary
Base score
2.9
Impact score
2.9
Exploitability score
5.5
Vector string
AV:A/AC:M/Au:N/C:N/I:P/A:N

Weaknesses

nvd@nist.gov
CWE-287
ics-cert@hq.dhs.gov
CWE-294

Social media

Hype score
Not currently trending

Configurations