CVE-2018-4845

Published Jun 26, 2018

Last updated 4 days ago

Overview

Description
A vulnerability has been identified in RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems (All versions_without_ use of Siemens Healthineers Informatics products), RAPIDLab 1200 Series (All versions < V3.3 _with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions >= V3.0 _with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (V2.4.X_with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions =< V2.3 _with_ Siemens Healthineers Informatics products), RAPIDPoint 400 systems (All versions _with_ Siemens Healthineers Informatics products). Remote attackers with either local or remote credentialed access to the "Remote View" feature might be able to elevate their privileges, compromising confidentiality, integrity, and availability of the system. No special skills or user interaction are required to perform this attack. At the time of advisory publication, no public exploitation of this security vulnerability is known. Siemens Healthineers confirms the security vulnerability and provides mitigations to resolve the security issue.
Source
productcert@siemens.com
NVD status
Modified

Risk scores

CVSS 3.1

Type
Primary
Base score
8.8
Impact score
5.9
Exploitability score
2.8
Vector string
CVSS:3.1/AV:N/AC:L/PR:L/UI:N/S:U/C:H/I:H/A:H
Severity
HIGH

CVSS 2.0

Type
Primary
Base score
6.5
Impact score
6.4
Exploitability score
8
Vector string
AV:N/AC:L/Au:S/C:P/I:P/A:P

Weaknesses

productcert@siemens.com
CWE-284
nvd@nist.gov
CWE-269

Social media

Hype score
Not currently trending

Configurations