CVE-2019-18241

Published Nov 26, 2019

Last updated 5 years ago

Overview

Description
In Philips IntelliBridge EC40 and EC80, IntelliBridge EC40 Hub all versions, and IntelliBridge EC80 Hub all versions, the SSH server running on the affected products is configured to allow weak ciphers. This could enable an unauthorized attacker with access to the network to capture and replay the session and gain unauthorized access to the EC40/80 hub.
Source
ics-cert@hq.dhs.gov
NVD status
Analyzed

Risk scores

CVSS 3.1

Type
Primary
Base score
6.5
Impact score
3.6
Exploitability score
2.8
Vector string
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
Severity
MEDIUM

CVSS 2.0

Type
Primary
Base score
3.3
Impact score
2.9
Exploitability score
6.5
Vector string
AV:A/AC:L/Au:N/C:P/I:N/A:N

Weaknesses

nvd@nist.gov
CWE-326
ics-cert@hq.dhs.gov
CWE-326

Social media

Hype score
Not currently trending

Configurations