CVE-2019-5024
Published Apr 11, 2019
Last updated 2 years ago
Overview
- Description
- A restricted environment escape vulnerability exists in the “kiosk mode” function of Capsule Technologies SmartLinx Neuron 2 medical information collection devices running versions 9.0.3 or lower. A specific series of keyboard inputs can escape the restricted environment, resulting in full administrator access to the underlying operating system. An attacker can connect to the device via USB port with a keyboard or other HID device to trigger this vulnerability.
- Source
- talos-cna@cisco.com
- NVD status
- Analyzed
Social media
- Hype score
- Not currently trending
Risk scores
CVSS 3.1
- Type
- Primary
- Base score
- 7.6
- Impact score
- 6
- Exploitability score
- 0.9
- Vector string
- CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
- Severity
- HIGH
CVSS 3.0
- Type
- Secondary
- Base score
- 7.6
- Impact score
- 6
- Exploitability score
- 0.9
- Vector string
- CVSS:3.0/AV:P/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
- Severity
- HIGH
CVSS 2.0
- Type
- Primary
- Base score
- 7.2
- Impact score
- 10
- Exploitability score
- 3.9
- Vector string
- AV:L/AC:L/Au:N/C:C/I:C/A:C
Weaknesses
- nvd@nist.gov
- NVD-CWE-noinfo
- talos-cna@cisco.com
- CWE-693
Configurations
[ { "nodes": [ { "negate": false, "cpeMatch": [ { "criteria": "cpe:2.3:o:capsuletech:smartlinx_neuron_2_firmware:*:*:*:*:*:*:*:*", "vulnerable": true, "matchCriteriaId": "0786573E-89D0-4D3F-A1A7-3AAD1102C092", "versionEndIncluding": "9.0.3" } ], "operator": "OR" }, { "negate": false, "cpeMatch": [ { "criteria": "cpe:2.3:h:capsuletech:smartlinx_neuron_2:-:*:*:*:*:*:*:*", "vulnerable": false, "matchCriteriaId": "A1ECBBAB-ADE3-4F31-97D5-56B818DD4276" } ], "operator": "OR" } ], "operator": "AND" } ]