CVE-2019-6540

Published Mar 26, 2019

Last updated 3 years ago

Overview

Description
The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement encryption. An attacker with adjacent short-range access to a target product can listen to communications, including the transmission of sensitive data.
Source
ics-cert@hq.dhs.gov
NVD status
Analyzed

Social media

Hype score
Not currently trending

Risk scores

CVSS 3.1

Type
Primary
Base score
6.5
Impact score
3.6
Exploitability score
2.8
Vector string
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
Severity
MEDIUM

CVSS 2.0

Type
Primary
Base score
3.3
Impact score
2.9
Exploitability score
6.5
Vector string
AV:A/AC:L/Au:N/C:P/I:N/A:N

Weaknesses

nvd@nist.gov
CWE-319
ics-cert@hq.dhs.gov
CWE-319

Configurations