- Description
- Sigma Spectrum Infusion System v's6.x (model 35700BAX) and Baxter Spectrum Infusion System Version(s) 8.x (model 35700BAX2) at the application layer uses an unauthenticated clear-text communication channel to send and receive system status and operational data. This could allow an attacker that has circumvented network security measures to view sensitive non-private data or to perform a man-in-the-middle attack.
- Source
- ics-cert@hq.dhs.gov
- NVD status
- Modified
CVSS 3.1
- Type
- Primary
- Base score
- 9.8
- Impact score
- 5.9
- Exploitability score
- 3.9
- Vector string
- CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
- Severity
- CRITICAL
CVSS 2.0
- Type
- Primary
- Base score
- 5
- Impact score
- 2.9
- Exploitability score
- 10
- Vector string
- AV:N/AC:L/Au:N/C:P/I:N/A:N
- Hype score
- Not currently trending
[
{
"nodes": [
{
"negate": false,
"cpeMatch": [
{
"criteria": "cpe:2.3:o:baxter:sigma_spectrum_infusion_system_firmware:*:*:*:*:*:*:*:*",
"vulnerable": true,
"matchCriteriaId": "0FCC49AB-B0AD-44FF-ACEB-3C26EA4F2C3C",
"versionEndIncluding": "6.05",
"versionStartIncluding": "6.0"
},
{
"criteria": "cpe:2.3:o:baxter:sigma_spectrum_infusion_system_firmware:8.0:*:*:*:*:*:*:*",
"vulnerable": true,
"matchCriteriaId": "72D0F7CB-3D24-4A8D-826D-ACB20ACBEB1C"
}
],
"operator": "OR"
},
{
"negate": false,
"cpeMatch": [
{
"criteria": "cpe:2.3:h:baxter:sigma_spectrum_infusion_system:-:*:*:*:*:*:*:*",
"vulnerable": false,
"matchCriteriaId": "49E25260-EC14-4E98-A86B-CBBE47E26AE5"
}
],
"operator": "OR"
}
],
"operator": "AND"
}
]