CVE-2020-12040

Published Jun 29, 2020

Last updated 4 years ago

Overview

Description
Sigma Spectrum Infusion System v's6.x (model 35700BAX) and Baxter Spectrum Infusion System Version(s) 8.x (model 35700BAX2) at the application layer uses an unauthenticated clear-text communication channel to send and receive system status and operational data. This could allow an attacker that has circumvented network security measures to view sensitive non-private data or to perform a man-in-the-middle attack.
Source
ics-cert@hq.dhs.gov
NVD status
Analyzed

Risk scores

CVSS 3.1

Type
Primary
Base score
9.8
Impact score
5.9
Exploitability score
3.9
Vector string
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Severity
CRITICAL

CVSS 2.0

Type
Primary
Base score
5
Impact score
2.9
Exploitability score
10
Vector string
AV:N/AC:L/Au:N/C:P/I:N/A:N

Weaknesses

nvd@nist.gov
CWE-319
ics-cert@hq.dhs.gov
CWE-319

Social media

Hype score
Not currently trending

Configurations