Overview
- Description
- An issue was discovered on Dr Trust ECG Pen 2.00.08 devices. Because the Bluetooth LE support is implemented without a requirement for pairing or security, any attacker can access the GATT server of the device and can sniff the data being broadcasted while a measurement is being done. Also, saved data can also be extracted over a Bluetooth connection. In addition, an attacker can launch a man-in-the-middle attack against data integrity.
- Source
- cve@mitre.org
- NVD status
- Analyzed
Risk scores
CVSS 3.1
- Type
- Primary
- Base score
- 6.5
- Impact score
- 3.6
- Exploitability score
- 2.8
- Vector string
- CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
- Severity
- MEDIUM
CVSS 2.0
- Type
- Primary
- Base score
- 3.3
- Impact score
- 2.9
- Exploitability score
- 6.5
- Vector string
- AV:A/AC:L/Au:N/C:P/I:N/A:N
Weaknesses
- nvd@nist.gov
- NVD-CWE-noinfo
Social media
- Hype score
- Not currently trending
Configurations
[ { "nodes": [ { "negate": false, "cpeMatch": [ { "criteria": "cpe:2.3:o:drtrust:electrocardiogram_pen_firmware:2.00.08:*:*:*:*:*:*:*", "vulnerable": true, "matchCriteriaId": "DEC90897-AD8D-4072-A28E-6326302F64EE" } ], "operator": "OR" }, { "negate": false, "cpeMatch": [ { "criteria": "cpe:2.3:h:drtrust:electrocardiogram_pen:2.00.08:*:*:*:*:*:*:*", "vulnerable": false, "matchCriteriaId": "33AFB885-737E-419E-AB0B-905B7B54D700" } ], "operator": "OR" } ], "operator": "AND" } ]