Overview
- Description
- In Patient Information Center iX (PICiX) Versions C.02, C.03, PerformanceBridge Focal Point Version A.01, the product receives input that is expected to be well-formed (i.e., to comply with a certain syntax) but it does not validate or incorrectly validates that the input complies with the syntax, causing the certificate enrollment service to crash. It does not impact monitoring but prevents new devices from enrolling.
- Source
- ics-cert@hq.dhs.gov
- NVD status
- Modified
Risk scores
CVSS 3.1
- Type
- Primary
- Base score
- 4.3
- Impact score
- 1.4
- Exploitability score
- 2.8
- Vector string
- CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:L
- Severity
- MEDIUM
CVSS 2.0
- Type
- Primary
- Base score
- 3.3
- Impact score
- 2.9
- Exploitability score
- 6.5
- Vector string
- AV:A/AC:L/Au:N/C:N/I:N/A:P
Weaknesses
- ics-cert@hq.dhs.gov
- CWE-1286
Social media
- Hype score
- Not currently trending
Configurations
[ { "nodes": [ { "negate": false, "cpeMatch": [ { "criteria": "cpe:2.3:a:philips:patient_information_center_ix:b.02:*:*:*:*:*:*:*", "vulnerable": true, "matchCriteriaId": "C4FA1FEC-5139-48C1-856C-8062436AE6C6" }, { "criteria": "cpe:2.3:a:philips:patient_information_center_ix:c.02:*:*:*:*:*:*:*", "vulnerable": true, "matchCriteriaId": "F3220AB2-AC0D-4AC2-90D8-76C02FC693EB" }, { "criteria": "cpe:2.3:a:philips:patient_information_center_ix:c.03:*:*:*:*:*:*:*", "vulnerable": true, "matchCriteriaId": "AE1FB3E9-E269-434A-B1C2-D54C40F437BE" }, { "criteria": "cpe:2.3:a:philips:performancebridge_focal_point:a.01:*:*:*:*:*:*:*", "vulnerable": true, "matchCriteriaId": "3B4CF59B-32DC-4F48-88C5-77B96E937E93" } ], "operator": "OR" } ] } ]