- Description
- A vulnerability has been identified in DCA Vantage Analyzer (All versions < V4.5 are affected by CVE-2020-7590. In addition, serial numbers < 40000 running software V4.4.0 are also affected by CVE-2020-15797). Affected devices use a hard-coded password to protect the onboard database. This could allow an attacker to read and or modify the onboard database. Successful exploitation requires direct physical access to the device.
- Source
- productcert@siemens.com
- NVD status
- Analyzed
CVSS 3.1
- Type
- Primary
- Base score
- 6.8
- Impact score
- 5.9
- Exploitability score
- 0.9
- Vector string
- CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
- Severity
- MEDIUM
CVSS 2.0
- Type
- Primary
- Base score
- 4.6
- Impact score
- 6.4
- Exploitability score
- 3.9
- Vector string
- AV:L/AC:L/Au:N/C:P/I:P/A:P
- productcert@siemens.com
- CWE-259
- Hype score
- Not currently trending
[
{
"nodes": [
{
"negate": false,
"cpeMatch": [
{
"criteria": "cpe:2.3:h:siemens:dca_vantage_analyzer:-:*:*:*:*:*:*:*",
"vulnerable": false,
"matchCriteriaId": "EDE5221F-8629-4DC4-AD8A-93EED671469B"
}
],
"operator": "OR"
},
{
"negate": false,
"cpeMatch": [
{
"criteria": "cpe:2.3:o:siemens:dca_vantage_analyzer_firmware:*:*:*:*:*:*:*:*",
"vulnerable": true,
"matchCriteriaId": "2B4F34A3-AB7B-4528-9F27-3073E25A5176",
"versionEndExcluding": "4.5.0.0"
}
],
"operator": "OR"
}
],
"operator": "AND"
}
]