CVE-2021-37166

Published Aug 2, 2021

Last updated a year ago

Overview

Description
A buffer overflow issue leading to denial of service was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. When HMI3 starts up, it binds a local service to a TCP port on all interfaces of the device, and takes extensive time for the GUI to connect to the TCP socket, allowing the connection to be hijacked by an external attacker.
Source
cve@mitre.org
NVD status
Modified

Risk scores

CVSS 3.1

Type
Primary
Base score
7.5
Impact score
3.6
Exploitability score
3.9
Vector string